FDA Device recall Z-2726-2024
Stryker iBur 4.0mm Diamond Round, Distal Bend- intended to cut bone in the following manner: drilling, reaming, decorticating, shaping, dis…
- FDA Device
- Class II
- completed
- Published 2024-09-04
On 2024-09-04 the FDA classified a Class II recall of Stryker iBur 4.0mm Diamond Round, Distal Bend- intended to cut bone in the following manner: drilling, reaming, decorticating, shaping, dis… by Stryker, citing potential for the product to exhibit temperatures higher than specification where the bur shank meets the distal bushing. Temperatures higher than specification may lead to the po….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2726-2024 |
| Classification | Class II |
| Status | completed |
| Initiated | 2024-07-24 |
| Published | 2024-09-04 |
| Last updated | 2026-09-13 11:57 UTC |
| Company | Stryker |
| Distribution | US |
Product
Stryker iBur 4.0mm Diamond Round, Distal Bend- intended to cut bone in the following manner: drilling, reaming, decorticating, shaping, dissecting, shaving, and smoothing for the following medical applications: Neuro; Spine; Ear, Nose, and Throat (ENT)/ Otorhinolaryngology; and Endoscopic applications Ref: 8431-012-040D
Reason
Potential for the product to exhibit temperatures higher than specification where the bur shank meets the distal bushing. Temperatures higher than specification may lead to the potential for minor tissue/structure damage or tissue/structure damage from thermal injury which may need medical/surgical intervention.
Identifiers
| code_info | GTIN: 07613327501278 Lot Numbers: 22334017 22339017 23004017 23016017 23085017 23115017 23195017 23299017 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: completed.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2726-2024%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.