FDA Device recall Z-2723-2014
Reusable laparoscopic instruments are packaged as non-sterile
- FDA Device
- Class II
- terminated
- Published 2014-10-08
On 2014-10-08 the FDA classified a Class II recall of Reusable laparoscopic instruments are packaged as non-sterile by Instrumed International, citing additional caution statements have been added to the Instructions For Use for monopolar electrodes. 1) To avoid tissue carbonation, the operation voltage of the HF generator must….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2723-2014 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2014-05-30 |
| Published | 2014-10-08 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Instrumed International |
| Distribution | US |
Product
Reusable laparoscopic instruments are packaged as non-sterile. Cleaning and sterilization of devices must occur prior to use. The instruments are intended to be used by qualified physicians to provide access, illumination and visualization of internal structures and for manipulating soft tissue (grasping, cutting, coagulating, dissecting and suturing) in a wide variety of diagnostic and therapeutic laparoscopic/ urologic closed and minimally invasive procedures. It is inserted via a trocar sleeve and is used to prepare, coagulate and cut tissue.
Reason
Additional caution statements have been added to the Instructions For Use for monopolar electrodes. 1) To avoid tissue carbonation, the operation voltage of the HF generator must not exceed 650 peak voltage (Vp) for all monopolar electrodes. and 2) For all monopolar electrodes, the useful life of this product is less than or equal to 50 cycles and less than or equal to 2 years.
Identifiers
| code_info | 1) Instrumed Item No.: 250-08680; Boss Instruments Catalog Code: 35-8015; Lot Number: 030411, 100110, 110210; 2) Inst…1) Instrumed Item No.: 250-08680; Boss Instruments Catalog Code: 35-8015; Lot Number: 030411, 100110, 110210; 2) Instrumed Item No.: 250-08700; Boss Instruments Catalog Code: 35-8016; Lot Number: 030111; 3) Instrumed Item No.: 250-08700; Boss Instruments Catalog Code: 35-8016; Lot Number: 30111; 4) Instrumed Item No.: 250-08760; Boss Instruments Catalog Code: 35-8020; Lot Number: 020411, 070111, 100412, 110412; 5) Instrumed Item No.: 250-08770; Boss Instruments Catalog Code: 35-8020L; Lot Number: 020112, 110211, 120310; 6) Instrumed Item No.: 250-08800; Boss Instruments Catalog Code: 35-8025; Lot Number: 060110, 100112; 7) Instrumed Item No.: 250-08820; Boss Instruments Catalog Code: 35-8026 ; Lot Number: 032811; 8) Instrumed Item No.: 250-08860; Boss Instruments Catalog Code: 35-8030; Lot Number: 050413, 070412, 110510, 090110; 9) Instrumed Item No.: 250-08880; Boss Instruments Catalog Code: 35-8030L; Lot Number: 090110; 10) Instrumed Item No.: 250-08890; Boss Instr |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2723-2014%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.