Transonic Flow Probe, 3 mm, Catalog Number HQC3FMC -- Product Usage: to measure flow intra-operatively

On 2015-09-09 the FDA classified a Class II recall of Transonic Flow Probe, 3 mm, Catalog Number HQC3FMC -- Product Usage: to measure flow intra-operatively by Transonic Systems, citing software defect; programming errors in the blood flow probes EPROM. Transonic received several complaints from users indicating that when the flow probe is connected to a flow met….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2721-2015
ClassificationClass II
Statusterminated
Initiated2015-07-27
Published2015-09-09
Last updated2026-09-13 11:53 UTC
CompanyTransonic Systems
DistributionUS

Product

Transonic Flow Probe, 3 mm, Catalog Number HQC3FMC -- Product Usage: to measure flow intra-operatively.

Reason

Software defect; programming errors in the blood flow probes EPROM. Transonic received several complaints from users indicating that when the flow probe is connected to a flow meter the meter reads Used Probe or Irrelevant Use Number.

Identifiers

code_info
Serial Numbers: HQC3FMC2501, HQC3FMC2502, HQC3FMC2503, HQC3FMC2504, HQC3FMC2505, HQC3FMC2506, HQC3FMC2507, HQC3FMC2508…Serial Numbers: HQC3FMC2501, HQC3FMC2502, HQC3FMC2503, HQC3FMC2504, HQC3FMC2505, HQC3FMC2506, HQC3FMC2507, HQC3FMC2508, HQC3FMC2509, HQC3FMC2510, HQC3FMC2511, HQC3FMC2512, HQC3FMC2513, HQC3FMC2514, HQC3FMC2515, HQC3FMC2516, HQC3FMC2517, HQC3FMC2518, HQC3FMC2519, HQC3FMC2520, HQC3FMC2521, HQC3FMC2522, HQC3FMC2523, HQC3FMC2524, HQC3FMC2525, HQC3FMC2526, HQC3FMC2527, HQC3FMC2528, HQC3FMC2529, HQC3FMC2530, HQC3FMC2531, HQC3FMC2532, HQC3FMC2533, HQC3FMC2534, HQC3FMC2535, HQC3FMC2636

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2721-2015%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.