FDA Device recall Z-2699-2020

regard Migrating Indicator, STEAM Migrating Indicator Strips, REORDER # 335576REG - Product Usage: The integrating indicator is designed to…

On 2020-08-05 the FDA classified a Class II recall of regard Migrating Indicator, STEAM Migrating Indicator Strips, REORDER # 335576REG - Product Usage: The integrating indicator is designed to… by Resource Optimization And Innovation, citing the firm received reports from customer that the Regard Migrating Indicator shown malfunctions and does not go into the "accept" window at the time of use.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2699-2020
ClassificationClass II
Statusterminated
Initiated2020-06-17
Published2020-08-05
Last updated2026-09-13 11:55 UTC
CompanyResource Optimization And Innovation
DistributionUS

Product

regard Migrating Indicator, STEAM Migrating Indicator Strips, REORDER # 335576REG - Product Usage: The integrating indicator is designed to chemically react over time with the critical parameters of steam sterilization cycle within a specified tolerance.

Reason

The firm received reports from customer that the Regard Migrating Indicator shown malfunctions and does not go into the "accept" window at the time of use.

Identifiers

code_infoLot 719

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.