FDA Device recall Z-2670-2026
Radius VSM ECG pre-connected Set, REF: 4842, 4695
- FDA Device
- Class II
- ongoing
- Published 2026-07-08
On 2026-07-08 the FDA classified a Class II recall of Radius VSM ECG pre-connected Set, REF: 4842, 4695 by Masimo, citing two devices recalled: 1) Electrocardiogram (ECG) electrode assembly ECG feature may incorrectly detect extreme bradycardia, tachycardia and trigger alarms on normal heart rate pat….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2670-2026 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2026-02-20 |
| Published | 2026-07-08 |
| Last updated | 2026-09-13 11:58 UTC |
| Company | Masimo |
| Distribution | US |
Product
Radius VSM ECG pre-connected Set, REF: 4842, 4695
Reason
Two devices recalled: 1) Electrocardiogram (ECG) electrode assembly ECG feature may incorrectly detect extreme bradycardia, tachycardia and trigger alarms on normal heart rate patients, which may distract clinicians away from management of true positive alarms, which could delay treatment; 2) Small and medium size blood pressure cuffs may contain rough edges, may result in skin irritation/redness.
Identifiers
| code_info | REF/UDI-DI/Lot: 4842/10843997011003/25AJH, 25KMA, 25MSC 4695/10843997016015/24N17, 25BBJ, 25DMH, 25FBR, 24KLV, 24N17…REF/UDI-DI/Lot: 4842/10843997011003/25AJH, 25KMA, 25MSC 4695/10843997016015/24N17, 25BBJ, 25DMH, 25FBR, 24KLV, 24N17, 25AGZ, 25BBJ, 25D10, 25DMH, 25FBR, 25GJM, 25KBF |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:58 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2670-2026%22