Jelco, Hypodermic needle Pro¿ needles 19G X1 1/2" 800/CA Model No. EL 1915 & 19G X 2" 600/CA Model No. SB 5021, Rx Only, Sterile EO

On 2017-07-12 the FDA classified a Class II recall of Jelco, Hypodermic needle Pro¿ needles 19G X1 1/2" 800/CA Model No. EL 1915 & 19G X 2" 600/CA Model No. SB 5021, Rx Only, Sterile EO by Smiths Medical Asd, citing smiths Medical became aware on January 11, 2017, that syringe assemblies provided by Nipro Medical Corporation contained pin holes in the side wall of the hub. The defect was not….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2670-2017
ClassificationClass II
Statusterminated
Initiated2017-04-04
Published2017-07-12
Last updated2026-09-13 11:54 UTC
CompanySmiths Medical Asd
DistributionIN

Product

Jelco, Hypodermic needle Pro¿ needles 19G X1 1/2" 800/CA Model No. EL 1915 & 19G X 2" 600/CA Model No. SB 5021, Rx Only, Sterile EO.

Reason

Smiths Medical became aware on January 11, 2017, that syringe assemblies provided by Nipro Medical Corporation contained pin holes in the side wall of the hub. The defect was not detected during Smiths Medical incoming inspection. The syringe assemblies were used to manufacture seven (7) finished good lots of Hypodermic Needle Pro¿ needles, Smiths Medical part numbers EL 1915 and SB5021.

Identifiers

code_info3297460,3297461 Model No. EL1915. 3297458,3297459,3297465, 3297466,3297467 Model No. SB5021.