RecallRadar

FDA Device recall Z-2665-2014

Siemens CentraLink Data Management System, software version 14

On 2014-09-24 the FDA classified a Class II recall of Siemens CentraLink Data Management System, software version 14 by Siemens Healthcare Diagnostics, citing siemens Healthcare Diagnostics has identified a task scheduler issue with the CentraLink" Data Management System, software version 14. The issue may cause the software to sto….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2665-2014
ClassificationClass II
Statusterminated
Initiated2014-08-19
Published2014-09-24
Last updated2026-09-13 11:53 UTC
CompanySiemens Healthcare Diagnostics
DistributionUS

Product

Siemens CentraLink Data Management System, software version 14. The CentraLink system software is a network solution provider and multi-system data manager for the instruments and lab automation systems (LAS) within the lab.

Reason

Siemens Healthcare Diagnostics has identified a task scheduler issue with the CentraLink" Data Management System, software version 14. The issue may cause the software to stop executing commands, including uploading validated results to the LIS. The issue is related to an internal software timer that overflows after 24 days. Automated rules triggered by task schedulers or by actions in the CentraLink user interface may also be affected. As each customer configuration is unique, in addition to the delay in uploading of validated results, the specific consequences of discontinued command execution will vary.

Identifiers

code_infoCatalog Numbers: 10810210, 10811596, 10814296, 10814298, 10814877, 10815474

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2665-2014%22