RecallRadar

FDA Device recall Z-2659-2025

MAGNETOM Prisma

On 2025-10-08 the FDA classified a Class I recall of MAGNETOM Prisma by Siemens Medical Solutions Usa, citing there is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the desi….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2659-2025
ClassificationClass I
Statusongoing
Initiated2025-08-28
Published2025-10-08
Last updated2026-09-13 11:57 UTC
CompanySiemens Medical Solutions Usa
DistributionUS

Product

MAGNETOM Prisma. Model Number: 10849582.

Reason

There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.

Identifiers

code_info
Model Number: 10849582. UDI Numbers: (01)04056869006727(21)166324, (01)04056869006727(21)66104, (01)04056869006727(21)1…Model Number: 10849582. UDI Numbers: (01)04056869006727(21)166324, (01)04056869006727(21)66104, (01)04056869006727(21)166097, (01)04056869006727(21)166024, (01)04056869006727(21)166018, (01)04056869006727(21)66109, (01)04056869006727(21)66111, (01)04056869006727(21)166114, (01)04056869006727(21)166242, (01)04056869006727(21)166268, (01)04056869006727(21)166200, (01)04056869006727(21)166047, (01)04056869006727(21)66047, (01)04056869006727(21)166292, (01)04056869006727(21)166041, (01)04056869006727(21)166007, (01)04056869006727(21)166037, (01)04056869006727(21)66112, (01)04056869006727(21)166025, (01)04056869006727(21)166003, (01)04056869006727(21)66102, (01)04056869006727(21)166254, (01)04056869006727(21)166074, (01)04056869006727(21)166281, (01)04056869006727(21)166220, (01)04056869006727(21)66027, (01)04056869006727(21)166221, (01)04056869006727(21)166321, (01)04056869006727(21)166193, (01)04056869006727(21)166012, (01)04056869006727(21)166121, (01)04056869006727(21)66053, (01)0405686

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2659-2025%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.