RecallRadar

FDA Device recall Z-2658-2023

Heartmate 3 LVAS Implant Kit, REF: 106524US, 106524INT

On 2023-10-11 the FDA classified a Class II recall of Heartmate 3 LVAS Implant Kit, REF: 106524US, 106524INT by Thoratec, citing inability to start and/or complete the coring procedure due to inadequate sharpness of the Apical Coring Knife, provided with Left Ventricular Assist System, could result in an ex….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2658-2023
ClassificationClass II
Statusongoing
Initiated2023-08-07
Published2023-10-11
Last updated2026-09-13 11:56 UTC
CompanyThoratec
DistributionUS

Product

Heartmate 3 LVAS Implant Kit, REF: 106524US, 106524INT; Heartmate XVE & HeartMate II Apical Coring Knife, REF: 1050

Reason

Inability to start and/or complete the coring procedure due to inadequate sharpness of the Apical Coring Knife, provided with Left Ventricular Assist System, could result in an extended procedure or surgical time while the backup coring knife is acquired.

Identifiers

code_info
REF/UDI-DI/Serial Numbers: 106524US/00813024013297/MLP-034137 to MLP-034744, MLP-035485 to MLP-035714, MLP-036330 to ML…REF/UDI-DI/Serial Numbers: 106524US/00813024013297/MLP-034137 to MLP-034744, MLP-035485 to MLP-035714, MLP-036330 to MLP-037708, MLP-038104 to MLP-038999, MLP-039000 to MLP-039999, MLP-040000 to MLP-040999, MLP-041000 to MLP-041621; 106524INT/00813024011712 (OUS)/MLP-032482, MLP-035542 to MLP-035667, MLP-037218 to MLP-037527, MLP-038008 to MLP-038989, MLP-039014 to MLP-039945, MLP-040009 to MLP-040896, MLP-041013 to MLP-041589; 1050/00813024010227: 135021-135025, 134241-134246, 134248, 134250, 134251, 134240, 135060, 135061, 135062, 134247, 132849-132853, 134239; Apical Coring Knife Distributed Separately Serial Number Range: 132849 to 135062

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2658-2023%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.