RecallRadar

FDA Device recall Z-2657-2025

MAGNETOM Connectom.X. Model Number: 11371480

On 2025-10-08 the FDA classified a Class I recall of MAGNETOM Connectom.X. Model Number: 11371480 by Siemens Medical Solutions Usa, citing there is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the desi….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2657-2025
ClassificationClass I
Statusongoing
Initiated2025-08-28
Published2025-10-08
Last updated2026-09-13 11:57 UTC
CompanySiemens Medical Solutions Usa
DistributionUS

Product

MAGNETOM Connectom.X. Model Number: 11371480.

Reason

There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.

Identifiers

code_infoModel Number: 11371480. UDI Numbers: N/A. Serial Numbers: 237551.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2657-2025%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.