FDA Device recall Z-2655-2020

Safe-T 8 Fr Thora-Para Tray With 8 Fr. Catheter and Sharps Injury Protection - Drug Free

On 2020-07-29 the FDA classified a Class II recall of Safe-T 8 Fr Thora-Para Tray With 8 Fr. Catheter and Sharps Injury Protection - Drug Free by Carefusion 2200, citing the Universal Drainage Set (UDS) may contain valves that leak internally which may result in the introduction of up to 30cc of air into the patient.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2655-2020
ClassificationClass II
Statusterminated
Initiated2020-05-28
Published2020-07-29
Last updated2026-09-13 11:55 UTC
CompanyCarefusion 2200
DistributionAL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY

Product

Safe-T 8 Fr Thora-Para Tray With 8 Fr. Catheter and Sharps Injury Protection - Drug Free

Reason

The Universal Drainage Set (UDS) may contain valves that leak internally which may result in the introduction of up to 30cc of air into the patient.

Identifiers

code_infoCatalog: TPT1000SDF UDI: 50885403108493; Batches: 0001320515, 0001321729, 0001323273, 0001323703, 0001332819, 0001334405, 0001338453, 0001344142, 0001344740