FDA Device recall Z-2641-2020
Computed Tomography X-ray System, Model: uCT 550, GTIN: 06971576831036 Intended use The uCT Computed Tomography X-ray System uCT 530/uCT 55…
- FDA Device
- Class II
- ongoing
- Published 2020-07-29
On 2020-07-29 the FDA classified a Class II recall of Computed Tomography X-ray System, Model: uCT 550, GTIN: 06971576831036 Intended use The uCT Computed Tomography X-ray System uCT 530/uCT 55… by Shanghai United Imaging Healthcare, citing two issues were identified with the computed tomography x-ray system including a service function which may cause false marking of a bad channel resulting in ring artifacts, and p….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2641-2020 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2020-05-23 |
| Published | 2020-07-29 |
| Last updated | 2026-09-13 11:55 UTC |
| Company | Shanghai United Imaging Healthcare |
| Distribution | TX |
Product
Computed Tomography X-ray System, Model: uCT 550, GTIN: 06971576831036 Intended use The uCT Computed Tomography X-ray System uCT 530/uCT 550 is a computed tomography X-ray system intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes and indicated for the whole body (including head, neck and vascular).
Reason
Two issues were identified with the computed tomography x-ray system including a service function which may cause false marking of a bad channel resulting in ring artifacts, and potential intermittent scout scanning interruption due to occasional angle signal drift.. If these problems occur, it may be necessary to rescan the patient resulting in an additional dose of radiation and the possible need for additional contrast medium.
Identifiers
| code_info | Serial Numbers: 353020, 353021, and 353023 with software version R001.11.0.729741-Re-1100 |
|---|