FDA Device recall Z-2635-2023
Ivenix Infusion System (IIS), Large Volume Pump (LVP) Software: 5.0.0 and 5.0.1 Part Number: LVP-SW-0004
- FDA Device
- Class II
- ongoing
- Published 2023-10-04
On 2023-10-04 the FDA classified a Class II recall of Ivenix Infusion System (IIS), Large Volume Pump (LVP) Software: 5.0.0 and 5.0.1 Part Number: LVP-SW-0004 by Fresenius Kabi Usa, citing start-up time for the flow rate range greater than 50 ml/hr, and less than 200 ml/hr may exceed the specified 6 minutes, depending on the characteristics of air bubble entrapment….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2635-2023 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2022-04-22 |
| Published | 2023-10-04 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Fresenius Kabi Usa |
| Distribution | US |
Product
Ivenix Infusion System (IIS), Large Volume Pump (LVP) Software: 5.0.0 and 5.0.1 Part Number: LVP-SW-0004. Indicated for use in a hospital and in outpatient care environments for the controlled administration of fluids through clinically accepted routes of administration.
Reason
Start-up time for the flow rate range greater than 50 ml/hr, and less than 200 ml/hr may exceed the specified 6 minutes, depending on the characteristics of air bubble entrapment within the Intermediate Pumping Chamber (IPC) of the administration set
Identifiers
| code_info | (UDI): 00811505030122 Software Version: 5.0.0 and 5.0.1 |
|---|