Ivenix Infusion System (IIS), Large Volume Pump (LVP) Software: 5.0.0 and 5.0.1 Part Number: LVP-SW-0004

On 2023-10-04 the FDA classified a Class II recall of Ivenix Infusion System (IIS), Large Volume Pump (LVP) Software: 5.0.0 and 5.0.1 Part Number: LVP-SW-0004 by Fresenius Kabi Usa, citing start-up time for the flow rate range greater than 50 ml/hr, and less than 200 ml/hr may exceed the specified 6 minutes, depending on the characteristics of air bubble entrapment….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2635-2023
ClassificationClass II
Statusongoing
Initiated2022-04-22
Published2023-10-04
Last updated2026-09-13 11:56 UTC
CompanyFresenius Kabi Usa
DistributionUS

Product

Ivenix Infusion System (IIS), Large Volume Pump (LVP) Software: 5.0.0 and 5.0.1 Part Number: LVP-SW-0004. Indicated for use in a hospital and in outpatient care environments for the controlled administration of fluids through clinically accepted routes of administration.

Reason

Start-up time for the flow rate range greater than 50 ml/hr, and less than 200 ml/hr may exceed the specified 6 minutes, depending on the characteristics of air bubble entrapment within the Intermediate Pumping Chamber (IPC) of the administration set

Identifiers

code_info(UDI): 00811505030122 Software Version: 5.0.0 and 5.0.1