FDA Device recall Z-2631-2017
Pointe Scientific CRP(HS) Wide Range Control Set and Pointe Scientific CRP(HS) Wide Range Standard Set 5 Levels of 1 x 2 ml ready to use ca…
- FDA Device
- Class III
- terminated
- Published 2017-07-05
On 2017-07-05 the FDA classified a Class III recall of Pointe Scientific CRP(HS) Wide Range Control Set and Pointe Scientific CRP(HS) Wide Range Standard Set 5 Levels of 1 x 2 ml ready to use ca… by Medtest Holdings, citing individual vials of calibrator and standard were mislabeled; specifically the units on the vial labels are incorrectly listed as mg/dl when the correct unit is mg/L.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2631-2017 |
| Classification | Class III |
| Status | terminated |
| Initiated | 2009-07-21 |
| Published | 2017-07-05 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Medtest Holdings |
| Distribution | CA, FL, HI, MD, MS, NJ, NV, OK, PA |
Product
Pointe Scientific CRP(HS) Wide Range Control Set and Pointe Scientific CRP(HS) Wide Range Standard Set 5 Levels of 1 x 2 ml ready to use calibrator or standard The Pointe Scientific CRP (HS) Wide Range Multi-Calibrator Set is intended to be used for the calibration of the manufacturer's CRP (High Sensitivity) Wide Range immunoturbimidmetric assay. For in vitro diagnostic use only. The Pointe Scientific CRP (HS) Wide Range Multi-Standard Set Is intended to be used for the calibration of the manufacturer's CRP (High Sensitivity) Wide Range immunoturbimidmetric assay. For in vitro diagnostic use only
Reason
Individual vials of calibrator and standard were mislabeled; specifically the units on the vial labels are incorrectly listed as mg/dl when the correct unit is mg/L.
Identifiers
| code_info | Model #'s C7568-CTL and C7568-STD Lots: C7568-STD: 717303-131. Exp 6/2009. C7568-CTL: 902701-111 and 902701-131. Exp 1/2011. |
|---|