FDA Device recall Z-2621-2016
Selective Salpingography Catheter with Beacon Tip Cannula, Manipulator/ Injector, Uterine Used for injection of contrast medium into the fa…
- FDA Device
- Class II
- terminated
- Published 2016-08-31
On 2016-08-31 the FDA classified a Class II recall of Selective Salpingography Catheter with Beacon Tip Cannula, Manipulator/ Injector, Uterine Used for injection of contrast medium into the fa… by Cook, citing increase in reports of polymer degradation of the catheter tip, resulting in tip fracture and/or separation. degradation could include loss of device function, separation of a dev….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2621-2016 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2016-04-15 |
| Published | 2016-08-31 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Cook |
| Distribution | US |
Product
Selective Salpingography Catheter with Beacon Tip Cannula, Manipulator/ Injector, Uterine Used for injection of contrast medium into the fallopian tube(s) for selective salpingography.
Reason
Increase in reports of polymer degradation of the catheter tip, resulting in tip fracture and/or separation. degradation could include loss of device function, separation of a device segment leading to medical intervention, or complications resulting from a separated segment. Such complications include device fragments in the vascular system, genitourinary system, or other soft tissues. Fragments within the vascular system could result in embolization to the heart or lungs, or occluding blood flow to end organs.
Identifiers
| code_info | All Lots. Catalog/product number; Global/order number J-SSG-504000 G17868 |
|---|