RecallRadar

FDA Device recall Z-2616-2016

Haskal Transjugular Intrahepatic Portal Access Set Introducer, catheter Intended for transjugular liver access in diagnostic and interventi…

On 2016-08-31 the FDA classified a Class II recall of Haskal Transjugular Intrahepatic Portal Access Set Introducer, catheter Intended for transjugular liver access in diagnostic and interventi… by Cook, citing increase in reports of polymer degradation of the catheter tip, resulting in tip fracture and/or separation. degradation could include loss of device function, separation of a dev….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2616-2016
ClassificationClass II
Statusterminated
Initiated2016-04-15
Published2016-08-31
Last updated2026-09-13 11:54 UTC
CompanyCook
DistributionUS

Product

Haskal Transjugular Intrahepatic Portal Access Set Introducer, catheter Intended for transjugular liver access in diagnostic and interventional procedures.

Reason

Increase in reports of polymer degradation of the catheter tip, resulting in tip fracture and/or separation. degradation could include loss of device function, separation of a device segment leading to medical intervention, or complications resulting from a separated segment. Such complications include device fragments in the vascular system, genitourinary system, or other soft tissues. Fragments within the vascular system could result in embolization to the heart or lungs, or occluding blood flow to end organs.

Identifiers

code_infoAll Lots. Catalog/product number; Global/order number HTPS-100 G36126

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2616-2016%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.