RecallRadar

FDA Device recall Z-2615-2026

Reprocessed Electrophysiology Catheters and Ultrasound Catheters Item no. 401206RH, St. Jude Medical Response" Diagnostic Electrophysiology…

On 2026-07-08 the FDA classified a Class I recall of Reprocessed Electrophysiology Catheters and Ultrasound Catheters Item no. 401206RH, St. Jude Medical Response" Diagnostic Electrophysiology… by Medline Industries, citing expansion of RES 98277. Specific lots of reprocessed catheters may contain small particles of residual material. If devices with residual particulates on patient-contacting surfac….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2615-2026
ClassificationClass I
Statusongoing
Initiated2026-05-15
Published2026-07-08
Last updated2026-09-13 11:58 UTC
CompanyMedline Industries
DistributionUS

Product

Reprocessed Electrophysiology Catheters and Ultrasound Catheters Item no. 401206RH, St. Jude Medical Response" Diagnostic Electrophysiology Catheter 4 Elec. 10mm; Item no. 401305RH, St. Jude Medical Response" Diagnostic Electrophysiology Catheter 10 Elec. 2-5-2mm.

Reason

Expansion of RES 98277. Specific lots of reprocessed catheters may contain small particles of residual material. If devices with residual particulates on patient-contacting surfaces are used, there is a potential risk of an inflammatory response or systemic infection. If particulate were to dislodge during use and contact circulating blood or the heart, it may produce a granulomatous reaction or thrombus formation, which may lead to cerebral or pulmonary embolism, or deep vein thrombosis.

Identifiers

code_infoLot EP250203 Item no. 401206RH - UDI-DI 10197344043767; Item no. 401305RH - UDI-DI 10197344043637.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:58 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2615-2026%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.