RecallRadar

FDA Device recall Z-2613-2016

Shuttle Select Slip-Cath Catheter, Intravascular, Diagnostic The catheters are intended for use in angiographic procedures by physicians tr…

On 2016-08-31 the FDA classified a Class II recall of Shuttle Select Slip-Cath Catheter, Intravascular, Diagnostic The catheters are intended for use in angiographic procedures by physicians tr… by Cook, citing increase in reports of polymer degradation of the catheter tip, resulting in tip fracture and/or separation. degradation could include loss of device function, separation of a dev….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2613-2016
ClassificationClass II
Statusterminated
Initiated2016-04-15
Published2016-08-31
Last updated2026-09-13 11:54 UTC
CompanyCook
DistributionUS

Product

Shuttle Select Slip-Cath Catheter, Intravascular, Diagnostic The catheters are intended for use in angiographic procedures by physicians trained and experienced in angiographic techniques

Reason

Increase in reports of polymer degradation of the catheter tip, resulting in tip fracture and/or separation. degradation could include loss of device function, separation of a device segment leading to medical intervention, or complications resulting from a separated segment. Such complications include device fragments in the vascular system, genitourinary system, or other soft tissues. Fragments within the vascular system could result in embolization to the heart or lungs, or occluding blood flow to end organs.

Identifiers

code_info
All Lots. Catalog prefixes/suffixes SCBR5.5-/-SHTL SCBR6.5-/-SHTL Catalog/product number; Global/order number SCB…All Lots. Catalog prefixes/suffixes SCBR5.5-/-SHTL SCBR6.5-/-SHTL Catalog/product number; Global/order number SCBR5.5-35-125-P-NS-H1-SHTL G31129 SCBR5.5-35-125-P-NS-JB1-SHTL G31130 SCBR5.5-35-125-P-NS-JB2-SHTL G36361 SCBR5.5-35-125-P-NS-MPA-SHTL G23697 SCBR5.5-35-125-P-NS-SIM1-SHTL G23699 SCBR5.5-35-125-P-NS-SIM2-SHTL G36360 SCBR5.5-35-125-P-NS-VERT-SHTL G23698 SCBR5.5-35-125-P-NS-VTK-SHTL G36359 SCBR5.5-35-135-P-NS-ANG-SHTL G48154 SCBR5.5-35-150-P-NS-0-SHTL G23705 SCBR5.5-35-150-P-NS-ANG-SHTL G23704 SCBR5.5-35-75-P-NS-ANG-SHTL G48152 SCBR6.5-35-125-P-NS-H1-SHTL G31134 SCBR6.5-35-125-P-NS-JB1-SHTL G31135 SCBR6.5-35-125-P-NS-JB2-SHTL G36364 SCBR6.5-35-125-P-NS-SIM1-SHTL G23701 SCBR6.5-35-125-P-NS-SIM2-SHTL G36363 SCBR6.5-35-125-P-N

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2613-2016%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.