On 2026-07-08 the FDA classified a Class I recall of Omnipod (Eros) Pod by Insulet, citing external soft cannula damage during manufacturing that results in insulin leaking around the Pod instead of being delivered to the user regardless of basal or bolus delivery. The….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2605-2026 |
| Classification | Class I |
| Status | ongoing |
| Initiated | 2026-05-20 |
| Published | 2026-07-08 |
| Last updated | 2026-09-13 11:58 UTC |
| Company | Insulet |
| Distribution | US |
Product
Omnipod (Eros) Pod. Includes the below Model/REF Numbers: 1. POD-ZXP425 (10 Pack), POD- ZXP420 (Single Pod). 2. ZXR425 (10-Pack), ZXR420 (Single Pod). 3. ZXP425 (10 Pack), ZXP420 (Single Pod).
Reason
External soft cannula damage during manufacturing that results in insulin leaking around the Pod instead of being delivered to the user regardless of basal or bolus delivery. The primary failure mode is pump under-delivery due to loss of insulin to an external leak. If insulin is not delivered properly, users may experience high blood glucose levels due to under-delivery of insulin. In the most severe cases, prolonged and persistent high blood glucose levels can lead to diabetic ketoacidosis (DKA), hyperglycemic hyperosmolar syndrome (HHS), or even death.
Identifiers
| code_info | 1. Model/REF Number: POD-ZXP425 (10 Pack), POD- ZXP420 (Single Pod). Pack UDI-DI: 20385081120033. Single Pod UDI-DI: 10…1. Model/REF Number: POD-ZXP425 (10 Pack), POD- ZXP420 (Single Pod). Pack UDI-DI: 20385081120033. Single Pod UDI-DI: 10385081120036. Lot Numbers: L72514. 2. Model/REF Number: ZXR425 (10-Pack), ZXR420 (Single Pod). Pack UDI-DI: N/A. Single Pod UDI-DI: N/A. Lot Numbers: L72516, L72521. 3. Model/REF Number: ZXP425 (10 Pack), ZXP420 (Single Pod). Pack UDI-DI: 20385081120033. Single Pod UDI-DI: 10385081120036. Lot Numbers: L72515, L72509, L72513, L72511, L72510, L71480, L72514. |
|---|