FDA Device recall Z-2604-2025
Philips CT systems labeled as the following: 1. CT 6000, Model Number: 728307
- FDA Device
- Class II
- ongoing
- Published 2025-10-01
On 2025-10-01 the FDA classified a Class II recall of Philips CT systems labeled as the following: 1. CT 6000, Model Number: 728307 by Philips North America, citing the patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2604-2025 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2025-08-13 |
| Published | 2025-10-01 |
| Last updated | 2026-09-13 11:57 UTC |
| Company | Philips North America |
| Distribution | US |
Product
Philips CT systems labeled as the following: 1. CT 6000, Model Number: 728307.
Reason
The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.
Identifiers
| code_info | 1. Model Number: 728307; UDI-DI: 00884838104600; Serial Numbers: 87150, 87169, 87118, 87168, 87166, 87164. |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.