Flow sensors that may be installed in the following GEHC anesthesia machines: (1) Aestiva MRI, GTIN 0080682102339
- FDA Device
- Class II
- ongoing
- Published 2021-10-06
On 2021-10-06 the FDA classified a Class II recall of Flow sensors that may be installed in the following GEHC anesthesia machines: (1) Aestiva MRI, GTIN 0080682102339 by Ge Healthcare, citing the flow sensors built prior to June 2021 could have damaged tubes with small punctures or cuts that could cause leaks resulting in incorrect anesthesia machine tidal volumes pote….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2596-2021 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2021-08-02 |
| Published | 2021-10-06 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Ge Healthcare |
| Distribution | US |
Product
Flow sensors that may be installed in the following GEHC anesthesia machines: (1) Aestiva MRI, GTIN 0080682102339; (2) Aestiva 7900; (3) Aestiva 7100; (4) Aespire System Carestation, S/5 Protiva Config., Aespire 100 Compact Wall Mount (Aespire 7100); (5) S/5 Aespire 7900; (6) Aespire View; (7) Avance; (8) Avance CS2, GTIN 00840682102292; (9) Aisys; (10) Aisys CS2, GTIN 00840682102322; (11) Amingo; (12) 9100 NXT Config, Aelite NXT; (13) Protiva 7100;
Reason
The flow sensors built prior to June 2021 could have damaged tubes with small punctures or cuts that could cause leaks resulting in incorrect anesthesia machine tidal volumes potentially leading to over-delivery of volume to the patient.
Identifiers
| code_info | The flow sensors built prior to June 2021 that are installed in the devices/anesthesia machines. |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2596-2021%22