RecallRadar

FDA Device recall Z-2592-2023

Membrane Changer Single-Use, REF MC, each package contains 3 changers

On 2023-09-27 the FDA classified a Class II recall of Membrane Changer Single-Use, REF MC, each package contains 3 changers by Sentec Ag, citing during rework of components during the manufacturing process, an additional piece of material may have been left in the device, which may result in an impact on measurement perfor….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2592-2023
ClassificationClass II
Statusongoing
Initiated2023-07-25
Published2023-09-27
Last updated2026-09-13 11:56 UTC
CompanySentec Ag
DistributionUS

Product

Membrane Changer Single-Use, REF MC, each package contains 3 changers; Membrane Changer Reloadable, REF MC-R, each package contains 3 changers; and Membrane Changer Insert, REF MC-I, each package contains 5 inserts. All are components of the Sentec Digital Monitoring System (SDMS).

Reason

During rework of components during the manufacturing process, an additional piece of material may have been left in the device, which may result in an impact on measurement performance.

Identifiers

code_info
REF MC: 230441, 230224, 230209, 230191, 230185, 230167, 230165, 230146, 230125, 230016, 225261, 225226, 225183, 225173…REF MC: 230441, 230224, 230209, 230191, 230185, 230167, 230165, 230146, 230125, 230016, 225261, 225226, 225183, 225173, 225149, 225147, 225126, 225122, 225088, 225073, 225063, 225006; UDI-DI numbers 07640121880933, 17640121880930; REF MC-R: 230461, 230329, 230351, 230235, 230083, 230053, 230043, 230008, 230005, 225253, 225221; UDI-DI numbers 07640121880087, 17640121880084, 27640121880081; REF MC-I: 230372, 230339, 230336, 230229, 230206, 230132, 230117, 230051, 230044, 230012, 225299, 225298, 225280, 225255, 225218, 225216, 225190, 225185, 225184, 225175, 225120, 225091, 225089, 225064, 225058, 225033, 225023, 221064, 221056; UDI-DI numbers 07640121880506, 17640121880503. REVISED 9/21/2023: Additional lot number added for the U.S. for REF MC-I - 225119

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2592-2023%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.