FDA Device recall Z-2591-2018

ENVOY 500 ISE CALIBRATOR KIT, reference 55117 (contained 6 x 20 mL Calibrator High level and 6 x 20 mL Calibrator Low level)

On 2018-08-08 the FDA classified a Class II recall of ENVOY 500 ISE CALIBRATOR KIT, reference 55117 (contained 6 x 20 mL Calibrator High level and 6 x 20 mL Calibrator Low level) by Elitech Clinical Systems Sas, citing ELITech Clinical Systems SAS initiated this recall because some users of ENVOY 500 ISE CALIBRATOR Kit (Part number: 55117), for Envoy 500 systems are observing trouble while calib….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2591-2018
ClassificationClass II
Statusterminated
Initiated2017-04-18
Published2018-08-08
Last updated2026-09-13 11:54 UTC
CompanyElitech Clinical Systems Sas
Distributionn/a

Product

ENVOY 500 ISE CALIBRATOR KIT, reference 55117 (contained 6 x 20 mL Calibrator High level and 6 x 20 mL Calibrator Low level). Model/Catalog Number: 55117

Reason

ELITech Clinical Systems SAS initiated this recall because some users of ENVOY 500 ISE CALIBRATOR Kit (Part number: 55117), for Envoy 500 systems are observing trouble while calibrating with some vials. Users contacted the firm when they received the messages "ISE OUT OF REPRODUCIBILITY" or "ISE SLOPE OUT OF RANGE" (i.e., failed calibration). This has resulted in delay in obtaining patient results until the calibrator lot is replaced.

Identifiers

code_infoLot number 7038 , Manufacture date 03/2017, Expiration date 02/2019