FDA Device recall Z-2591-2018
ENVOY 500 ISE CALIBRATOR KIT, reference 55117 (contained 6 x 20 mL Calibrator High level and 6 x 20 mL Calibrator Low level)
- FDA Device
- Class II
- terminated
- Published 2018-08-08
On 2018-08-08 the FDA classified a Class II recall of ENVOY 500 ISE CALIBRATOR KIT, reference 55117 (contained 6 x 20 mL Calibrator High level and 6 x 20 mL Calibrator Low level) by Elitech Clinical Systems Sas, citing ELITech Clinical Systems SAS initiated this recall because some users of ENVOY 500 ISE CALIBRATOR Kit (Part number: 55117), for Envoy 500 systems are observing trouble while calib….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2591-2018 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2017-04-18 |
| Published | 2018-08-08 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Elitech Clinical Systems Sas |
| Distribution | n/a |
Product
ENVOY 500 ISE CALIBRATOR KIT, reference 55117 (contained 6 x 20 mL Calibrator High level and 6 x 20 mL Calibrator Low level). Model/Catalog Number: 55117
Reason
ELITech Clinical Systems SAS initiated this recall because some users of ENVOY 500 ISE CALIBRATOR Kit (Part number: 55117), for Envoy 500 systems are observing trouble while calibrating with some vials. Users contacted the firm when they received the messages "ISE OUT OF REPRODUCIBILITY" or "ISE SLOPE OUT OF RANGE" (i.e., failed calibration). This has resulted in delay in obtaining patient results until the calibrator lot is replaced.
Identifiers
| code_info | Lot number 7038 , Manufacture date 03/2017, Expiration date 02/2019 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2591-2018%22