FDA Device recall Z-2590-2024
PORTEX Bivona Tracheostomy Silicone Tube, labeled as: a) NEONATAL TRACHEOSTOMY TUBE , Product Code/List Number/Item Code 60N025
- FDA Device
- Class I
- ongoing
- Published 2024-09-04
On 2024-09-04 the FDA classified a Class I recall of PORTEX Bivona Tracheostomy Silicone Tube, labeled as: a) NEONATAL TRACHEOSTOMY TUBE , Product Code/List Number/Item Code 60N025 by Smiths Medical Asd, citing smiths Medical has identified that the securement flange of specific lots of the Bivona Neonatal/Pediatric and Adult Tracheostomy products may tear because of a manufacturing defe….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2590-2024 |
| Classification | Class I |
| Status | ongoing |
| Initiated | 2024-05-29 |
| Published | 2024-09-04 |
| Last updated | 2026-09-13 11:57 UTC |
| Company | Smiths Medical Asd |
| Distribution | n/a |
Product
PORTEX Bivona Tracheostomy Silicone Tube, labeled as: a) NEONATAL TRACHEOSTOMY TUBE , Product Code/List Number/Item Code 60N025
Reason
Smiths Medical has identified that the securement flange of specific lots of the Bivona Neonatal/Pediatric and Adult Tracheostomy products may tear because of a manufacturing defect.
Identifiers
| code_info | a) UDI/DI 15021312005684, Product Code/List Number/Item Code 60N025, Lot Numbers: 3853476, 3868313, 3911441, 3952404,…a) UDI/DI 15021312005684, Product Code/List Number/Item Code 60N025, Lot Numbers: 3853476, 3868313, 3911441, 3952404, 3965318, 3991202, 4006287, 4006294, 4097838, 4125826, 4148013, 4185431, 4206833, 4221354, 4238779, 4331627, 4345039, 4348774, 4383653, 4400691 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2590-2024%22