Arjo Citadel beds originally assembled with the IndiGo modules during manufcturing, model number CX812A3J3AMAD0

On 2021-10-06 the FDA classified a Class II recall of Arjo Citadel beds originally assembled with the IndiGo modules during manufcturing, model number CX812A3J3AMAD0 by Arjohuntleigh Polska Sp Z O O, citing the IndiGo power cord might wear during use which could lead to its damage, resulting in the cord being cut, burnt or short circuit with limited emission of sparks.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2588-2021
ClassificationClass II
Statusterminated
Initiated2021-06-10
Published2021-10-06
Last updated2026-09-13 11:56 UTC
CompanyArjohuntleigh Polska Sp Z O O
DistributionUS

Product

Arjo Citadel beds originally assembled with the IndiGo modules during manufcturing, model number CX812A3J3AMAD0; and IndiGo modules (retrofit kits) that can be installed on Citadel (model no. INDIDAA) or Enterprise (model no. INDIJAA) beds manufactured without IndiGo modules assembled at the time of manufacturing and subsequently retrofitted on the beds.

Reason

The IndiGo power cord might wear during use which could lead to its damage, resulting in the cord being cut, burnt or short circuit with limited emission of sparks.

Identifiers

code_info
Manufacturing dates 5/25/2018-2/15/2021 (printed on both the bed and IndiGo module if the module was assembled on the b…Manufacturing dates 5/25/2018-2/15/2021 (printed on both the bed and IndiGo module if the module was assembled on the bed at the time of manufacture, or printed on the module if it was later installed on the bed.

Status history