FDA Device recall Z-2580-2018

Discovery MI, labeled as Discovery DMI 15 cm 3 ring, and Discovery DMI 20cm 4 ring

On 2018-08-08 the FDA classified a Class II recall of Discovery MI, labeled as Discovery DMI 15 cm 3 ring, and Discovery DMI 20cm 4 ring by Ge Healthcare, citing some CT and CT/PET systems recently installed with a GE supplied A! electrical panel may not be properly wired to the partial system UPS (Uninteruptible Power Supply).

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2580-2018
ClassificationClass II
Statusterminated
Initiated2018-05-24
Published2018-08-08
Last updated2026-09-13 11:54 UTC
CompanyGe Healthcare
DistributionUS

Product

Discovery MI, labeled as Discovery DMI 15 cm 3 ring, and Discovery DMI 20cm 4 ring

Reason

Some CT and CT/PET systems recently installed with a GE supplied A! electrical panel may not be properly wired to the partial system UPS (Uninteruptible Power Supply).

Identifiers

code_info
Model Number 5454001-170. Serial Number (System ID Number): CJRPX1800016CN ( 4550966PT11), CJRPX1800017CN (561422MI…Model Number 5454001-170. Serial Number (System ID Number): CJRPX1800016CN ( 4550966PT11), CJRPX1800017CN (561422MI), CJRPX1800018CN (617726PET1EB), CJRPX1800021CN (404616DMI)