FDA Device recall Z-2579-2026

Philips Telemetry Monitor 5500 1.4 GHz. Model Number: 867232

On 2026-07-01 the FDA classified a Class II recall of Philips Telemetry Monitor 5500 1.4 GHz. Model Number: 867232 by Philips North America, citing potential for device reset to default "NEW_DEVICE" state, which may lead to a to loss of device configuration and the equipment label, resulting in a failure to reconnect to the P….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2579-2026
ClassificationClass II
Statusongoing
Initiated2026-06-05
Published2026-07-01
Last updated2026-09-13 11:58 UTC
CompanyPhilips North America
DistributionUS

Product

Philips Telemetry Monitor 5500 1.4 GHz. Model Number: 867232. Includes the following system descriptions: TELE Monitor 5500, 1.4 Ghz, ECG only, ex; Telemetry Monitor 5500 1.4 GHz FAST EX.

Reason

Potential for device reset to default "NEW_DEVICE" state, which may lead to a to loss of device configuration and the equipment label, resulting in a failure to reconnect to the PIC iX central monitoring system.

Identifiers

code_info
Model Number: 867232. UDI-DI: (01)00884838102842. Software Version Number: 4.0.0.0. Serial Numbers: US001A002F, US001A0…Model Number: 867232. UDI-DI: (01)00884838102842. Software Version Number: 4.0.0.0. Serial Numbers: US001A002F, US001A00AW, US001A00C9, US001A00EW, US001A00F0, US001A00GM, US001A00N4, US001A00PB, US001A00PZ, US001A00R1, US001A00UE, US001A00VU, US001A00Z4, US001A013T, US001A018A, US001A018R.

Status history