FDA Device recall Z-2579-2014
PERSONA The Personalized Knee System PS Tibial Articular Surface Provisional Left with Size 3-5 CD Top
- FDA Device
- Class II
- terminated
- Published 2014-09-10
On 2014-09-10 the FDA classified a Class II recall of PERSONA The Personalized Knee System PS Tibial Articular Surface Provisional Left with Size 3-5 CD Top by Zimmer, citing reports of breakage of the Persona Knee Trial Articular Surface Provisionals (TASPs). The Persona primary knee surgical technique has been updated. This field action correction is….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2579-2014 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2014-08-07 |
| Published | 2014-09-10 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Zimmer |
| Distribution | US |
Product
PERSONA The Personalized Knee System PS Tibial Articular Surface Provisional Left with Size 3-5 CD Top; Size 6-9 CD Top; Size 3-5 EF Top; Size 6-9 EF Top; Size 10-11 EF Top; Size 6-9 GH Top; Size 10-12 GH Top; Size 10-12 J Top; and RIGHT Size 3-5 CD Top; Right Size 6-9 CD Top; Right Size 3-5 EF Top; Right Size 6-9 EF Top; Right Size 10-11 EF Top; Right Size 6-9 GH Top; Right Size 10-12 GH Top; and Right Size 10-12 J Top.
Reason
Reports of breakage of the Persona Knee Trial Articular Surface Provisionals (TASPs). The Persona primary knee surgical technique has been updated. This field action correction is to ensure that all distributors and users of this instrument are aware of the enhanced surgical techniques.
Identifiers
| code_info | Part Numbers: 42517400410, 42517400510, 42517400610, 42517400710, 42517400810, 42517400910, 42517401010, 42517401110, 4…Part Numbers: 42517400410, 42517400510, 42517400610, 42517400710, 42517400810, 42517400910, 42517401010, 42517401110, 42527400410, 42527400510, 42527400610, 42527400710, 42527400810, 42527400910, 42527401010, and 42527401110. Lot numbers:62043556, 62110045, 62121489, 62125589, 62147235, 62165546, 62168383, 62177247, 62311021, 62314722, 62386421, 62442838, 62525491, 62558585, 62605208, 62627292, 62683432, 62696865, 62717343, 62022344, 62041351, 62073566, 62104372, 62117744, 62121488, 62149175, 62187324, 62187325, 62247351, 62311022, 62314723, 62364961, 62391117, 62436468, 62474903, 62538403, 62564311, 62611089, 62669905, 62690895, 62720225, 62742767, 62747707, 62020946, 62028233, 62073570, 62110047, 62130976, 62187327, 62187329, 62189325, 62290822, 62305407, 62314724, 62386424, 62420258, 62469459, 62531662, 62551356, 62565041, 62584447, 62612266, 62687089, 62690049, 62690909, 62702763, 62724739, 62022345, 62024978, 62104374, 62124040, 62134006, 62200223, 62200224, 62233921, 62294595, 6 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2579-2014%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.