FDA Device recall Z-2577-2014
PERSONA The Personalized Knee System CR Tibial Articular Surface Provisional with the following: Left Size 3-9 CD Top
- FDA Device
- Class II
- terminated
- Published 2014-09-10
On 2014-09-10 the FDA classified a Class II recall of PERSONA The Personalized Knee System CR Tibial Articular Surface Provisional with the following: Left Size 3-9 CD Top by Zimmer, citing reports of breakage of the Persona Knee Trial Articular Surface Provisionals (TASPs). The Persona primary knee surgical technique has been updated. This field action correction is….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2577-2014 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2014-08-07 |
| Published | 2014-09-10 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Zimmer |
| Distribution | US |
Product
PERSONA The Personalized Knee System CR Tibial Articular Surface Provisional with the following: Left Size 3-9 CD Top; Left Size 3-11 EF Top; Left Size 7-12 GH Top; Left Size 9-12 J Top; Left Size 3-9 CD Top; Left Size 3-11 EF Top; Right Size 7-12 GH Top; and Right Size 9-12 J Top.
Reason
Reports of breakage of the Persona Knee Trial Articular Surface Provisionals (TASPs). The Persona primary knee surgical technique has been updated. This field action correction is to ensure that all distributors and users of this instrument are aware of the enhanced surgical techniques.
Identifiers
| code_info | Part Numbers: 42517000410, 42517000510, 42517000610, 42517000710, 42527000410, 42527000510, 42527000610, and 4252700071…Part Numbers: 42517000410, 42517000510, 42517000610, 42517000710, 42527000410, 42527000510, 42527000610, and 42527000710. Lot numbers: 62024975, 62043546, 62144792, 62152725, 62165538, 62187318, 62337961, 62372407, 62386776, 62397159, 62405917, 62456263, 62499170, 62551386, 62565051, 62568272, 62572572, 62663168, 62691384, 62707494, 62717346, 62020920, 62144793, 62156313, 62165539, 62184228, 62353932, 62356176, 62385100, 62420242, 62468277, 62538412, 62564340, 62611088, 62612263, 62690056, 62702295, 62717337, 62736078, 62024976, 62054776, 62147232, 62177243, 62187319, 62187320, 62359057, 62364955, 62391222, 62414006, 62461233, 62558583, 62564345, 62608889, 62638657, 62646286, 62702749, 62724348, 62394279, 62443397, 62487490, 62492802, 62506001, 62563915, 62568271, 62704633, 62055686, 62144796, 62161537, 62162597, 62256286, 62351297, 62356578, 62385099, 62404282, 62424802, 62455897, 62538392, 62565037, 62595855, 62611083, 62663162, 62680121, 62686205, 62695286, 62724347, 62724744, 6201 |
|---|