FDA Device recall Z-2575-2018

Liquid Cardiac Control - Level 3, cat. no

On 2018-08-08 the FDA classified a Class II recall of Liquid Cardiac Control - Level 3, cat. no by Randox Laboratories, citing the device is not suitable for the control of the Troponin T assay due to unacceptable variation between vials. The deviation in precision observed is up to 28%. IQC that is repor….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2575-2018
ClassificationClass II
Statusterminated
Initiated2018-06-08
Published2018-08-08
Last updated2026-09-13 11:54 UTC
CompanyRandox Laboratories
DistributionWV

Product

Liquid Cardiac Control - Level 3, cat. no. CQ5053 Product Usage: This product is intended for in vitro diagnostic use, in the quality control of Cardiac Markers on clinical chemistry and Immunoassay systems.

Reason

The device is not suitable for the control of the Troponin T assay due to unacceptable variation between vials. The deviation in precision observed is up to 28%. IQC that is reported as out of range could lead to a delay in reporting Troponin T results. If the failure occurs the operator may observe the quality control result for Troponin T falling outside range and/or increased imprecision for Quality Control monitoring.

Identifiers

code_infoLot 4245CK.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.