FDA Device recall Z-2568-2025
Automated Impella Controller (AIC) with the below brands and product codes
- FDA Device
- Class I
- ongoing
- Published 2025-10-01
On 2025-10-01 the FDA classified a Class I recall of Automated Impella Controller (AIC) with the below brands and product codes by Abiomed, citing specific Automated Impella Controllers (AICs) that have a Pump Driver Circuit Assembly do not meet current specifications. This may lead to decreased pump performance or pump stop….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2568-2025 |
| Classification | Class I |
| Status | ongoing |
| Initiated | 2025-08-20 |
| Published | 2025-10-01 |
| Last updated | 2026-09-13 11:57 UTC |
| Company | Abiomed |
| Distribution | AR, AZ, CA, CO, CT, FL, GA, ID, MA, MI, MO, MS, NJ, NY, OH, PA, TX |
Product
Automated Impella Controller (AIC) with the below brands and product codes. Used with heart assist pump. 1. Impella Controller, Packaged, US; Product Code: 0042-0000-US. 2. Impella Controller, Packaged, US, Loaner; Product Code: 0042-0000-US-L. 3. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA. 4. Impella Controller, Packaged, EU; Product Code: 0042-0000-EU. 5. Impella Controller, Packaged, UK; Product Code: 0042-0000-UK. 6. Impella Controller, UK, Loaner; Product Code: 0042-0000-UK-L.
Reason
Specific Automated Impella Controllers (AICs) that have a Pump Driver Circuit Assembly do not meet current specifications. This may lead to decreased pump performance or pump stop and trigger an Impella Failure or Impella Stopped Controller Failure alarm.
Identifiers
| code_info | 1. Product Code: 0042-0000-US; GTIN: 813502010022; Serial Numbers: IC1026, IC1029, IC1043, IC1046, IC1040, IC1047, IC10…1. Product Code: 0042-0000-US; GTIN: 813502010022; Serial Numbers: IC1026, IC1029, IC1043, IC1046, IC1040, IC1047, IC1088, IC1089, IC1093, IC1110, IC1167, IC1221, IC1225, IC1222, IC1258, IC1250, IC1272, IC1293, IC1320, IC1317, IC1353, IC1354, IC1355, IC1375, IC1377, IC1403, IC1446, IC1450, IC1452, IC1461, IC1496, IC1501, IC1508, IC1515, IC1516, IC1517, IC1518, IC1530, IC1534, IC1535, IC1539, IC1538, IC1531, IC1533, IC1540, IC1532, IC1551, IC1552, IC1550, IC1553, IC1646, IC1756, IC2255, IC2422, IC2489, IC3013. 2. Product Code: 0042-0000-US-L; GTIN: 813502010022; Serial Number: IC1281. 3. Product Code: 0042-0000-CA; GTIN: 813502011272; Serial Numbers: IC1543, IC1544. 4. Product Code: 0042-0000-EU; GTIN: 813502011289; Serial Numbers: IC1350, IC1383, IC1546, IC1548, IC1547, IC1549, IC1775, IC2243, IC4002. 5. Product Code: 0042-0000-UK; GTIN: 813502011296; Serial Numbers: IC1519, IC1542. 6. Product Code: 0042-0000-UK-L; GTIN: 813502011296; Serial Number: IC1493. |
|---|