RecallRadar

FDA Device recall Z-2563-2014

Roche cobas IT 1000 Product Usage: Intended Use: An electronic device intended to store, retrieve, and process laboratory data

On 2014-09-10 the FDA classified a Class II recall of Roche cobas IT 1000 Product Usage: Intended Use: An electronic device intended to store, retrieve, and process laboratory data by Roche Diagnostics Operations, citing potential that test results would be assigned to the wrong Patient ID if cobas IT 1000 receives an ADT ADMIT, TRANSFER or MERGE event message that contains a genuine Patient ID (e….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2563-2014
ClassificationClass II
Statusterminated
Initiated2014-07-31
Published2014-09-10
Last updated2026-09-13 11:53 UTC
CompanyRoche Diagnostics Operations
DistributionUS

Product

Roche cobas IT 1000 Product Usage: Intended Use: An electronic device intended to store, retrieve, and process laboratory data.

Reason

Potential that test results would be assigned to the wrong Patient ID if cobas IT 1000 receives an ADT ADMIT, TRANSFER or MERGE event message that contains a genuine Patient ID (e.g. 999999), which is the same as what has been configured as the Generic Patient ID (999999 by default).

Identifiers

code_info
Part Numbers: 04420306001 Serial #'s: 2706, 2724, 2743, 2744, 2745, 2749, 2750, 2751, 2752, 2754, 2755, 2764, 2770, 27…Part Numbers: 04420306001 Serial #'s: 2706, 2724, 2743, 2744, 2745, 2749, 2750, 2751, 2752, 2754, 2755, 2764, 2770, 2771, 2772, 2773, 2774, 2775, 2776, 2780, 2781, 2782, 2783, 2784, 2785, 2789, 2790, 2791, 2792, 2793, 2794, 2795, 2796, 2797, 2798, 2799, 2808, 2814, 2815, 2816, 2818, 2819, 2823, 2824, 2825, 2826, 2828, 2829, 2830, 2831, 2847, 2848, 2849, 2857, 2858, 2859, 2860, 2861, 2862, 2863, 2864, 2865, 2904, 2905, 2906, 2908, 2909, 2914, 2915, 2916, 2917, 2920, 2921, 2928, 2931, 2932, 2933, 2934, 2935, 2936, 2937, 2938, 2939, 2940, 2941, 2943, 2983, 2984, 2985, 2986, 2987, 2990, 2991, 2992, 2993, 2994, 2995, 2996, 2997, 2998, 3006, 3007, 3008, 3009, 3016, 3020, 3021, 3023, 3024, 3025, 3026, 3027, 3048, 3049, 3050, 3051, 3053, 3057, 3058, 3059, 3060, 3061, 3065, 3066, 3067, 3068, 3069, 3070, 3071, 3072, 3073, 3074, 3075, 3076, 3077, 3079, 3082, 3083, 3084, 3085, 3086, 3087, 3089, 3090, 3104, 3105, 3106, and 3107.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2563-2014%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.