FDA Device recall Z-2557-2018
cobas m 511 integrated hematology analyzer, Model Number 07261691190 Product Usage: The cobas m 511 integrated hematology analyzer is a qua…
- FDA Device
- Class II
- terminated
- Published 2018-08-01
On 2018-08-01 the FDA classified a Class II recall of cobas m 511 integrated hematology analyzer, Model Number 07261691190 Product Usage: The cobas m 511 integrated hematology analyzer is a qua… by Roche Diagnostics Hematology, citing discrepant results have been reported in patients with severe microcytic anemia (e.g.iron deficiency, thalassemia) and the following parameters are affected: RBC (red blood cell c….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2557-2018 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2018-05-07 |
| Published | 2018-08-01 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Roche Diagnostics Hematology |
| Distribution | n/a |
Product
cobas m 511 integrated hematology analyzer, Model Number 07261691190 Product Usage: The cobas m 511 integrated hematology analyzer is a quantitative, automated analyzer with cell locating capability. It is intended for in vitro diagnostic use by a skilled operator in the clinical laboratory. The system prepares a stained microscope slide from EDTA anticoagulated whole blood. It utilizes computer imaging to count the formed elements of blood and provide an image-based assessment of cell morphology, which may be reviewed by the operator, and also allows for manual classification of unclassified cells. The instrument reports the following parameters: RBC, HGB, HCT, MCV, MCH, MCHC, RDW, RDW-SD, %NRBC, #NRBC, WBC, %NEUT, #NEUT, %LYMPH, #LYMPH, %MONO, #MONO, %EO, #EO, %BASO, #BASO, PLT, MPV, %RET, #RET, HGB-RET.
Reason
Discrepant results have been reported in patients with severe microcytic anemia (e.g.iron deficiency, thalassemia) and the following parameters are affected: RBC (red blood cell count), HGB (hemoglobin concentration), MCH (mean corpuscular hemoglobin), HCT (hematocrit), and MCV (mean corpuscular volume).
Identifiers
| code_info | All systems with software version 1.0 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.