- Published 2017-06-21
On 2017-06-21 the FDA classified a Class II recall of Fem IM Nail 15mmdx44cm, Sterile, Rod, fixation, intramedullary and accessories- Specific Knee, Hip and Nail Implants by Zimmer Biomet, citing zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2553-2017 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2017-02-28 |
| Published | 2017-06-21 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Zimmer Biomet |
| Distribution | US |
Product
Fem IM Nail 15mmdx44cm, Sterile, Rod, fixation, intramedullary and accessories- Specific Knee, Hip and Nail Implants
Reason
Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.
Identifiers
| code_info | Item Number: 00225244015, Lot Number Expiry Date Before: March 2027 |
|---|