FDA Device recall Z-2548-2026

Hillrom VOLARA System P.CIRCUIT KIT (Single Patient Use Circuit), Model/Catalog Number REF M08085

On 2026-07-08 the FDA classified a Class I recall of Hillrom VOLARA System P.CIRCUIT KIT (Single Patient Use Circuit), Model/Catalog Number REF M08085 by Baxter Healthcare, citing there have been reports of air and medication leakage from the nebulizer cup during therapy of Volara system patient circuits. The leakage can lead to an oxygen level drop in the….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2548-2026
ClassificationClass I
Statusongoing
Initiated2026-05-21
Published2026-07-08
Last updated2026-09-13 11:58 UTC
CompanyBaxter Healthcare
DistributionUS

Product

Hillrom VOLARA System P.CIRCUIT KIT (Single Patient Use Circuit), Model/Catalog Number REF M08085; Oscillation Lung Expansion Therapy, positive pressure breathing device

Reason

There have been reports of air and medication leakage from the nebulizer cup during therapy of Volara system patient circuits. The leakage can lead to an oxygen level drop in the patient and ineffective nebulization, which may affect the delivery of the prescribed treatment. The issue has been associated with improper locking of the nebulizer cup after adding medication during user setup.

Identifiers

code_infoUDI: 00887761985018, Lot Numbers: All lot numbers distributed beginning on 4/28/2025

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:58 UTC. Current status: ongoing.