FDA Device recall Z-2539-2018
AD-TECH Medical Instrument Corporation Drill Sleeve Guides DSG-6-3-090-2.4N utilizing raw material (RM0677) from the following lots from ve…
- FDA Device
- Class II
- terminated
- Published 2018-08-01
On 2018-08-01 the FDA classified a Class II recall of AD-TECH Medical Instrument Corporation Drill Sleeve Guides DSG-6-3-090-2.4N utilizing raw material (RM0677) from the following lots from ve… by Ad Tech Medical Instrument, citing the diameter of the drill sleeve guide raw material was found to be under tolerance, potentially resulting in the drill bit seizing in the guide during surgery.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2539-2018 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2018-05-02 |
| Published | 2018-08-01 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Ad Tech Medical Instrument |
| Distribution | US |
Product
AD-TECH Medical Instrument Corporation Drill Sleeve Guides DSG-6-3-090-2.4N utilizing raw material (RM0677) from the following lots from vendor 00-0001062., AD0021502, AD0021628. Product Usage: The drill sleeve guides are accessories to the Ad-Tech Cranial Drill Bits. The Cranial Drill Bits and accessories are intended to be used for drilling holes in the skull for neurological procedures, such as brain biopsy, brain contacting electrode and electrode accessory device placement. The DSG-6.3-090-2.4N drill sleeve guides are intended to be used with the 2.4mm Cranial Drill Bit diameter.
Reason
The diameter of the drill sleeve guide raw material was found to be under tolerance, potentially resulting in the drill bit seizing in the guide during surgery.
Identifiers
| code_info | Catalog Number DSG-6.3-090-2.4N. Batch Number (Lot Number): 107104 (710220), 114390 (715001), 111423 (716180), 105176…Catalog Number DSG-6.3-090-2.4N. Batch Number (Lot Number): 107104 (710220), 114390 (715001), 111423 (716180), 105176 (613121), 113775 (719190), 115847 (712211), 113776 (719190), 115000 (714201), 114724 (711101), 116389 (712121), 117146 (810110). |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.