FDA Device recall Z-2536-2016
Stryker AutoPlex System
- FDA Device
- Class II
- terminated
- Published 2016-08-24
On 2016-08-24 the FDA classified a Class II recall of Stryker AutoPlex System by Stryker Instruments Div, citing the Piston Head can become separated from the Delivery Piston, blocking the injection assembly Valve resulting in a cement backflow towards the injector handle. Potential for dela….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2536-2016 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2016-06-14 |
| Published | 2016-08-24 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Stryker Instruments Div |
| Distribution | US |
Product
Stryker AutoPlex System. Model Numbers 0607-687-000, 0605-887-000, 0605-687-000, and 0605-683-000. For bone cement. Stryker AutoPlex System kits contain a cement mixer, cement injector, funnel, and an extension tube assembly whereby bone cement powder and liquid monomer are mixed inside a chamber and transferred into a cement injector automatically. The bone cement is then dispensed directly into a fracture site using the cement injector and introducer needle combination.
Reason
The Piston Head can become separated from the Delivery Piston, blocking the injection assembly Valve resulting in a cement backflow towards the injector handle. Potential for delay in surgery if additional cement needs to be prepared for the injection procedure.
Identifiers
| code_info | Model Number(s): 0607-687-000 0605-887-000 0605-687-000 0605-683-000 0607-687-000 Lot numbers: 16021012, 16027…Model Number(s): 0607-687-000 0605-887-000 0605-687-000 0605-683-000 0607-687-000 Lot numbers: 16021012, 16027022, 16049012, 16061012, 16092012, 16107012, 16022012, 16028012, 16049022, 16063012, 16098012, 16109012, 16025012, 16028022, 16049032, 16063022, 16100012, 16109022, 16025022, 16033012, 16053012, 16070012, 16104012, 16113022, 16026012, 16041012, 16056012, 16070022, 16104022, 16113032, 16026022, 16048012, 16056022, 16074012, 16105012, 16118012, 16027012, 16048022, 16060012, 16088012, 16106012, 16118022 0605-887-000 Lot numbers: 16041012, 16040012, 16050012, 16057012, 16069012, 16078012, 16104012, 16112012, 16124012 0605-687-000 Lot numbers: 16041012, 16047012, 16111012 0605-683-000 Lot numbers: 16071012 |
|---|