RecallRadar

FDA Device recall Z-2535-2018

HexaPOD evo RT Couch Top (also known as HexaPOD evo Module), a component of the HexaPOD evo RT system

On 2018-08-01 the FDA classified a Class II recall of HexaPOD evo RT Couch Top (also known as HexaPOD evo Module), a component of the HexaPOD evo RT system by Elekta, citing upon completion of a 6D workflow, the HexaPOD evo Module may be in a tilted state if it is moved to the position used for 3D (non-iGUIDE) treatments resulting in incorrect patient….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2535-2018
ClassificationClass II
Statusterminated
Initiated2018-05-16
Published2018-08-01
Last updated2026-09-13 11:54 UTC
CompanyElekta
DistributionUS

Product

HexaPOD evo RT Couch Top (also known as HexaPOD evo Module), a component of the HexaPOD evo RT system. Product Usage: To support and aid in positioning a patient during radiation therapy

Reason

Upon completion of a 6D workflow, the HexaPOD evo Module may be in a tilted state if it is moved to the position used for 3D (non-iGUIDE) treatments resulting in incorrect patient positioning.

Identifiers

code_info
UDI 04056719001698. Serial numbers: 42380, 42384, 42626, 03005, 42009, 42012, 42013, 42062, 42068, 42106, 42107, 421…UDI 04056719001698. Serial numbers: 42380, 42384, 42626, 03005, 42009, 42012, 42013, 42062, 42068, 42106, 42107, 42126, 42153, 42217, 42224, 42292, 42293, 42301, 42435, 42491, 42496, 42517, 42615, 42620, 42690, 42691, 42731, 03019, 42092, 42222, 42356, 42382, 42601, 42649, 42658, 42743, 42289, 03000, 42311, 42586, 42471, 03017, 03022, 42057, 42058, 42129, 42159, 42187, 42279, 02982, 02986, 02991, 42004, 42113, 42178, 42189, 42202, 42207, 42209, 42212, 42213, 42216, 42220, 42416, 42417, 42418, 42419, 42420, 42463, 42652, 42704, 42706, 02987, 03008, 03018, 42000, 42008, 42032, 42048, 42064, 42070, 42074, 42084, 42102, 42115, 42118, 42134, 42137, 42142, 42145, 42146, 42152, 42174, 42177, 42237, 42312, 42329, 42330, 42341, 42363, 42369, 42371, 42375, 42383, 42394, 42409, 42446, 42450, 42453, 42457, 42464, 42474, 42501, 42509, 42519, 42523, 42524, 42550, 42553, 42587, 42611, 42612, 42643, 42647, 42662, 42674, 42681, 42694, 42721, 42722, 42069, 03029, 42036, 42709, 42710, 42139, 42162, 42