FDA Device recall Z-2530-2023
Hamilton-T1, REF: MSP161339/02, ESM Shielding Set, Spare Part for ventilator
- FDA Device
- Class I
- ongoing
- Published 2023-09-27
On 2023-09-27 the FDA classified a Class I recall of Hamilton-T1, REF: MSP161339/02, ESM Shielding Set, Spare Part for ventilator by Hamilton Medical, citing degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to l….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2530-2023 |
| Classification | Class I |
| Status | ongoing |
| Initiated | 2023-07-26 |
| Published | 2023-09-27 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Hamilton Medical |
| Distribution | US |
Product
Hamilton-T1, REF: MSP161339/02, ESM Shielding Set, Spare Part for ventilator
Reason
Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation.
Identifiers
| code_info | UDI: 07630002800761 / Affected Serial Numbers: 4050193, 4050210, 4050194 and 4016669 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2530-2023%22