FDA Device recall Z-2530-2016
Perioscopy Sheath: Part number: 1008
- FDA Device
- Class II
- terminated
- Published 2016-08-24
On 2016-08-24 the FDA classified a Class II recall of Perioscopy Sheath: Part number: 1008 by Danville Materials, citing the Perioscopy Sheaths are currently provided with a label that indicates the Sheath has a 3 year shelf-life/expiry date. Danville Materials has become aware that validation data….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2530-2016 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2016-06-24 |
| Published | 2016-08-24 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Danville Materials |
| Distribution | n/a |
Product
Perioscopy Sheath: Part number: 1008; The Perioscopy Sheath is a single-use sterile component used with the Perioscopy System, which is a medical device that features miniaturized digital video technology, and enables the healthcare professional to see magnified details of the tooth anatomy and deposits below the gum line (i.e. calculus root fractures, restoration margins, tooth decay).
Reason
The Perioscopy Sheaths are currently provided with a label that indicates the Sheath has a 3 year shelf-life/expiry date. Danville Materials has become aware that validation data are not available to support the 3-year shelf-life/expiration dating claim.
Identifiers
| code_info | Part number 1008: Lot numbers: 1133; 1135; 1136; 1137; 1138; 1163; 1171; 1172; 1173; 1174; 1175; 1218; 1240; 1243;…Part number 1008: Lot numbers: 1133; 1135; 1136; 1137; 1138; 1163; 1171; 1172; 1173; 1174; 1175; 1218; 1240; 1243; 1258; 1264; 1264-1; 1275; 1276; 1277; 1311; 1314; 1315; 1325; 1326. |
|---|