Hamilton C1/T1/MR1, REF: MSP161502/12, Control Board Spare part for ventilator

On 2023-09-27 the FDA classified a Class I recall of Hamilton C1/T1/MR1, REF: MSP161502/12, Control Board Spare part for ventilator by Hamilton Medical, citing degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to l….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2529-2023
ClassificationClass I
Statusongoing
Initiated2023-07-26
Published2023-09-27
Last updated2026-09-13 11:56 UTC
CompanyHamilton Medical
DistributionUS

Product

Hamilton C1/T1/MR1, REF: MSP161502/12, Control Board Spare part for ventilator

Reason

Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation.

Identifiers

code_infoUDI: 07630002805247 / Affected Serial Numbers: 4001901, 4002139, 4003445, 4003567, 4003576, 4005595