FDA Device recall Z-2528-2023

Hamilton-MR1, REF: 161010, Intensive Care Ventilator

On 2023-09-27 the FDA classified a Class I recall of Hamilton-MR1, REF: 161010, Intensive Care Ventilator by Hamilton Medical, citing degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to l….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2528-2023
ClassificationClass I
Statusongoing
Initiated2023-07-26
Published2023-09-27
Last updated2026-09-13 11:56 UTC
CompanyHamilton Medical
DistributionUS

Product

Hamilton-MR1, REF: 161010, Intensive Care Ventilator

Reason

Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation.

Identifiers

code_infoUDI: 07630002800761 & 07630002813556 / Affected Serial Numbers: 3422 3423 3424 3425 3426 3427 3428 3429 3469 3503 3510 3800

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.