Stratus¿ CS STAT Fluorometric Analyzer-microprocessor-controlled instrument that measures certain analytes in body fluids for in vitro diag…

On 2016-08-17 the FDA classified a Class II recall of Stratus¿ CS STAT Fluorometric Analyzer-microprocessor-controlled instrument that measures certain analytes in body fluids for in vitro diag… by Siemens Healthcare Diagnostics, citing software defect, where either an Above Assay Range or an inaccurate value could potentially be reported, in the Stratus CS Acute Care Diagnostics System regarding TestPak calibrat….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2525-2016
ClassificationClass II
Statusterminated
Initiated2016-06-22
Published2016-08-17
Last updated2026-09-13 11:54 UTC
CompanySiemens Healthcare Diagnostics
Distributionn/a

Product

Stratus¿ CS STAT Fluorometric Analyzer-microprocessor-controlled instrument that measures certain analytes in body fluids for in vitro diagnostic use. Assays include ~hCG, CRP, CKMB, cTNI, D-Dimer, Myoglobin, and pBNP. SMN: 10444834, 10453531

Reason

Software defect, where either an Above Assay Range or an inaccurate value could potentially be reported, in the Stratus CS Acute Care Diagnostics System regarding TestPak calibration and quality control (QC) status when a TestPak is due to expire within 48 hours.

Identifiers

code_infoAll software versions and serial numbers

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.