FDA Device recall Z-2522-2020

uCT 760 System, Computed Tomography X-ray System

On 2020-07-15 the FDA classified a Class II recall of uCT 760 System, Computed Tomography X-ray System by Shanghai United Imaging Healthcare, citing two issues were identified with the computed tomography x-ray system including an intermittent issue which may affect image data leading to artifacts; and intermittent scout scann….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2522-2020
ClassificationClass II
Statusterminated
Initiated2020-05-08
Published2020-07-15
Last updated2026-09-13 11:55 UTC
CompanyShanghai United Imaging Healthcare
DistributionTX

Product

uCT 760 System, Computed Tomography X-ray System

Reason

Two issues were identified with the computed tomography x-ray system including an intermittent issue which may affect image data leading to artifacts; and intermittent scout scanning interruption due to occasional angle signal drift. This may result in the need for rescanning of the patient with the patient receiving an additional dose of radiation and possible use of additional contrast media.

Identifiers

code_info
Model uCT 760; Serial Numbers: 600229, 600230, 600232, 600271 with software version R001.11.0.729741-Re-1100, and seria…Model uCT 760; Serial Numbers: 600229, 600230, 600232, 600271 with software version R001.11.0.729741-Re-1100, and serial number 600282 with software version R001.11.0.785930-Re-1150.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.