FDA Device recall Z-2517-2015
KimVent* Multi-Access Port Replacement Catheter for Adults,Tracheostomy, 14 F (Product Code 2401453-5) A tracheobronchial suction catheter…
- FDA Device
- Class II
- terminated
- Published 2015-09-09
On 2015-09-09 the FDA classified a Class II recall of KimVent* Multi-Access Port Replacement Catheter for Adults,Tracheostomy, 14 F (Product Code 2401453-5) A tracheobronchial suction catheter… by Halyard Health, citing the thumb valve of some KIMVENT Closed Suction Systems remained partially open after suctioning was completed, even though the valve appeared to be closed or in the "LOCKED" posit….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2517-2015 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2015-07-29 |
| Published | 2015-09-09 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Halyard Health |
| Distribution | US |
Product
KimVent* Multi-Access Port Replacement Catheter for Adults,Tracheostomy, 14 F (Product Code 2401453-5) A tracheobronchial suction catheter is a device that is attached to a port of the airway circuit for use as needed to aspirate liquids or semisolids from a patient's upper airway.
Reason
The thumb valve of some KIMVENT Closed Suction Systems remained partially open after suctioning was completed, even though the valve appeared to be closed or in the "LOCKED" position.
Identifiers
| code_info | Product Code 2401453-5 - M5105T801, M5105T802 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2517-2015%22