FDA Device recall Z-2511-2024
IntelliVue Patient Monitor MX550 with L.x and M.x
- FDA Device
- Class II
- ongoing
- Published 2024-08-14
On 2024-08-14 the FDA classified a Class II recall of IntelliVue Patient Monitor MX550 with L.x and M.x by Philips North America, citing philips lntelliVue Patient Monitors MX400/450/500/550 Shipped or Upgraded with Incorrect Software Options. Changes were made to reflect some options becoming standard capabilities….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2511-2024 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2024-06-28 |
| Published | 2024-08-14 |
| Last updated | 2026-09-13 11:57 UTC |
| Company | Philips North America |
| Distribution | US |
Product
IntelliVue Patient Monitor MX550 with L.x and M.x; K.x provided with L.x entitlements Model Number: 866066. Indicated for use by health care professionals for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults, pediatrics, and neonates.
Reason
Philips lntelliVue Patient Monitors MX400/450/500/550 Shipped or Upgraded with Incorrect Software Options. Changes were made to reflect some options becoming standard capabilities for software-version N.x. In the process, the entitlements of software options MOS, M06, and M20 were removed for software versions L.x and M.x. As a result of these changes Patient Monitors manufactured with or updated to the latest versions of software L.x or M.x will not offer the capabilities> Mmonitors with software version K.x may experience this issue too, due to software version K being out of support and the devices being provided with L.x entitlements offered by software options MOS, M06, and M20. Potential for incorrect or delayed treatment for the patient,
Identifiers
| code_info | UDI: 00884838038783 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2511-2024%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.