Endo GIA 45mm Gray Articulating Vascular Reload with Tri-Staple Technology, Item Code EGIA45AV

On 2018-07-25 the FDA classified a Class II recall of Endo GIA 45mm Gray Articulating Vascular Reload with Tri-Staple Technology, Item Code EGIA45AV by Covidien Medtronic, citing device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading t….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2509-2018
ClassificationClass II
Statusongoing
Initiated2018-05-22
Published2018-07-25
Last updated2026-09-13 11:54 UTC
CompanyCovidien Medtronic
DistributionUS

Product

Endo GIA 45mm Gray Articulating Vascular Reload with Tri-Staple Technology, Item Code EGIA45AV

Reason

Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.

Identifiers

code_infoN7L0823KX, N8A0799KX & N8B1008KX