RecallRadar

FDA Device recall Z-2507-2018

Tri-Staple 2.0 Black Intelligent Reload for use with Signia stapling system 60 mm Extra Thick, Item Code SIG60AXT

On 2018-07-25 the FDA classified a Class II recall of Tri-Staple 2.0 Black Intelligent Reload for use with Signia stapling system 60 mm Extra Thick, Item Code SIG60AXT by Covidien Medtronic, citing device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading t….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2507-2018
ClassificationClass II
Statusongoing
Initiated2018-05-22
Published2018-07-25
Last updated2026-09-13 11:54 UTC
CompanyCovidien Medtronic
DistributionUS

Product

Tri-Staple 2.0 Black Intelligent Reload for use with Signia stapling system 60 mm Extra Thick, Item Code SIG60AXT

Reason

Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.

Identifiers

code_infoN7L0852X, N7M0130X, N8A0145X, N8A0336X, N7L0931X, N7M0639X, N8A0796X & N8B0993X

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2507-2018%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.