RecallRadar

FDA Device recall Z-2503-2018

Endo GIA 45 mm Articulating Medium/Thick Reload with Tri-Staple Technology, Item Code EGIA45AMT

On 2018-07-25 the FDA classified a Class II recall of Endo GIA 45 mm Articulating Medium/Thick Reload with Tri-Staple Technology, Item Code EGIA45AMT by Covidien Medtronic, citing device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading t….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2503-2018
ClassificationClass II
Statusongoing
Initiated2018-05-22
Published2018-07-25
Last updated2026-09-13 11:54 UTC
CompanyCovidien Medtronic
DistributionUS

Product

Endo GIA 45 mm Articulating Medium/Thick Reload with Tri-Staple Technology, Item Code EGIA45AMT

Reason

Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.

Identifiers

code_info
N7L0521KX, N7M0047KX, N7M0592KX, N7M1001KX, N7L0975KX, N7M0434KX, N7M0683KX, N8A0203KX, N7L1057KX, N7M0528KX, N7M0937K…N7L0521KX, N7M0047KX, N7M0592KX, N7M1001KX, N7L0975KX, N7M0434KX, N7M0683KX, N8A0203KX, N7L1057KX, N7M0528KX, N7M0937KX, N8A0677KX, N7L1098KX, N7M0529KX, N7M0949KX, N8C0144KX & N7L1099KX