FDA Device recall Z-2502-2018
VOLISTA StandOP Surgical Light, ARD568811961 VLT600DF AIM STP
- FDA Device
- Class II
- terminated
- Published 2018-07-25
On 2018-07-25 the FDA classified a Class II recall of VOLISTA StandOP Surgical Light, ARD568811961 VLT600DF AIM STP by Maquet Sas, citing certain VOLISTA StandOP Surgical Lights have been identified as having a potential light head detachment issue. Improper assembly of the bracket that connects the light head to th….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2502-2018 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2018-06-14 |
| Published | 2018-07-25 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Maquet Sas |
| Distribution | US |
Product
VOLISTA StandOP Surgical Light, ARD568811961 VLT600DF AIM STP; ARD568821961 VLT600DF AIM STP QL; ARD568811911 VLT600SF AIM STP; ARD568821911 VLT600SF AIM STP QL; ARD568821910 VLT600SF STP QL; ARD568811901 VLT600SF AIM STP Product Usage: The Maquet VOLISTA surgical lights are designed to illuminate the body of a patient during surgical operations, diagnostics and treatment. StandOP VOLISTA is a system of ceiling-mounted surgical lights suited for installation in surgical suites, examining rooms, doctors surgeries and out-patient consultations.
Reason
Certain VOLISTA StandOP Surgical Lights have been identified as having a potential light head detachment issue. Improper assembly of the bracket that connects the light head to the axle fork could result in the light head bracket breaking, causing the light head to detach from the arm and become suspended only from the cabling inside the arm. Use of an affected device may result in serious injury to patient and/or medical staff if the light head detaches and drops, impacting patient and/or medical staff during demonstration, service, or use.
Identifiers
| code_info | All Serial Numbers manufactured from March 10, 2014 until November 17, 2016 6. |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2502-2018%22